Strain parameters

Configure formulas and filters for strain measurements, including classic and vendor-specific modes.

Formula selection

  • strainView: new | geokon | smejkalka | pontex | true (legacy)
  • K: coefficient for legacy formula (used when strainView=true)
  • f0: Hz — reference frequency for legacy formula

Zero/reference and sensitivity

  • L0 (or Dig_0): digit reference at zero strain (f²/1000)
  • G: gauge factor (sensitivity)
  • gaugeLength: mm — physical gauge length used for deformation

Filtering

  • filterData: true/false — enable threshold filtering
  • filterParamL / filterParamH: lower/upper thresholds for strain values

Temperature channel scaling

  • tempMultiplier / tempDivider / tempOffset — adjust temperature channel from strain sensor

Optional displays

  • digitView: true/false — show digit values (f²/1000)
  • deformationView: true/false — show deformation in mm
  • frequencyView: true/false — show raw frequency in Hz
  • pascalView: true/false — convert to stress using Young’s modulus (see Unit data parameters for E)

Tips

  • Choose strainView first to match your sensor family
  • Set L0/Dig_0 and G accurately; verify with a known zero condition